STC «Promprilad»: Ukraine, 04080, Kyiv-080, P.B. 43 phone/fax: (044) 467-51-38, 467-51-39 EMail: ndt@ln.com.ua
Dynamic hardness gauge TDM-1 (release 1.5) is intended for high-speed measurements and hardness inspection of the structural, carbon and stainless steels and nonferrous metals made alloys, based on methods of Rockwell, Brinell, Vickers, Super-Rockwell, Shore and Leyb.
The dynamic hardness gauge is a compact device of TDM-2 type that is intended for express non destructive measurements and inspection of hardness of the structural, carbon steel and stainless steels and alloys made of nonferrous metals based on the Rockwell and Brinell scales.
Special-purpose eddy current flaw detector ВД-30НК-IVE is designed for detecting surface cracks in components made of ferromagnetic materials with rough-turned plate and curved surface
Densitometer for radiographic films entitled "Expert", is designed for measuring optical density of industrial radiographic films used during radiographic inspections of facilities.
Ultrasonic flaw detector UD3-71 is intended for defects detection, such as discontinuity and irregularities in materials, end items, semi-products and welded (soldered) joints; for the measurement of the flaw depth and coordinates; for the measurement of echo signals amplitude ratio; for the equivalent size measurement.
The flaw detector МД-4П (МД-4К) is designed for detecting surface flaws of discontinuity type in metals by means of the magnetic particle technique on the local sections of large-size components through originating stationary magnetic field applied.
Set of lenses with backlighting KLP-5 represents a power unit with a switch where the lenses with various magnification levels may be installed. The designed lens and the lamp are combined into one single device and may be easily replaced.
|
|
|
 |
|
Dynamic hardness gauge TDM-1 Dynamic hardness gauge TDM-1 (release 1.5) is intended for high-speed measurements and hardness inspection of the structural, carbon and stainless steels and nonferrous metals made alloys, based on methods of Rockwell, Brinell, Vickers, Super-Rockwell, Shore and Leyb at the laboratories, under shop and field conditions. In addition the device may be used to presorting materials depending on their elastic properties.
Specific features of the instrument: - calibration on references to measure strength of any non-porous materials (number of references required is 2…9) and internal calibration quality audit;
- automatic indications update depending on the transducer inclination angle entered to the instrument;
- "experiment" mode for laboratory measurements;
- single and multiple measurements mode;
- internal memory and PC input (Port RS232);
- battery supply and battery discharge level control;
- automatic flaw warning system;
- display backlighting;
- automatic translation of hardness number to the ultimate tension (in accordance with GOST 22761).
Performance specifications:
number of hardness scale | 12 |
|
measurement limits |
error |
resolution capability |
HRC Scale НВ Scale HV Scale HSD Scale |
20...70 90...450 375...900 30...100 |
±2.0 ±15 ±15 ±2.0 |
0.1 1.0 1.0 0.1 |
internal memory storage | 2000 measurements |
number of averaged single measurements | 3...99 |
operating temperature range | 10...40 °С |
dimensions | 157х84х30 mm |
weight | 0.4 kg | |
Set to be supplied:
Electronic unit | 1 |
Transducer with a cable | 1 |
Pusher | 1 |
In-built battery unit | 1 |
Charging unit | 1 |
Hardness gauge-1/computer connector | 1 |
Diskette with the software | 1 |
Spare indenter with a hard-alloy tag * | |
Leather casing * | 1 |
Certificate | 1 |
Transportation box | 1 | |
* the item supplied is to be charged extra
Dynamic hardness testers TDM-1, TDM-2
|
|
|
|
|
|
|
 |
|